Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med PM03: (1 year) CAL VER /5520 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 94 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Astro-Med PM03 Calibration Procedure Rev "A" # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: 20 min. - REQUIRED 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <75% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD INITIAL SETUP 1.008 DISP Press the ZERO DEFAULT. 1.008 DISP 1.008 DISP Ground Channels. 1.008 DISP 1.008 DISP Set for Bottom Baseline. 1.008 DISP 1.008 DISP Select Peak to Peak Mode. 1.009 HEAD {FULL SCALE AMPLIFIER ACCURACY} 1.010 HEAD FULL SCALE AMPLIFIER ACCURACY ({100 mV Range}) 1.011 DISP Set the UUT to the 100mV range. 1.012 5520 100mV S 2W 1.013 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 2.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({250 mV Range}) 2.002 DISP Set the UUT to the 250mV range. 2.003 5520 250mV S 2W 2.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 3.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({500 mV Range}) 3.002 DISP Set the UUT to the 500mV range. 3.003 5520 500mV S 2W 3.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 4.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({1 V Range}) 4.002 DISP Set the UUT to the 1V range. 4.003 5520 1V S 2W 4.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 5.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({5 V Range}) 5.002 DISP Set the UUT to the 5V range. 5.003 5520 5V S 2W 5.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 6.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({10 V Range}) 6.002 DISP Set the UUT to the 10V range. 6.003 5520 10V S 2W 6.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 7.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({50 V Range}) 7.002 DISP Set the UUT to the 50V range. 7.003 5520 50V S 2W 7.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 8.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({100 V Range}) 8.002 DISP Set the UUT to the 100V range. 8.003 5520 100V S 2W 8.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 9.001 HEAD FULL SCALE AMPLIFIER ACCURACY ({250 V Range}) 9.002 DISP Set the UUT to the 250V range. 9.003 5520 250V S 2W 9.004 EVAL Does {UUT indicate full scale deflection +/-1/2 div}? 10.001 HEAD {FREQUENCY RESPONSE} 10.002 DISP Set the trace to center scale. 10.002 DISP 10.002 DISP Set the UUT to the 10V range. 10.002 DISP 10.002 DISP Set to 50 division grid. 10.003 5520 10V 5kH SI S 2W 10.004 EVAL Is {display >35 division (-3dB point)}? 11.001 END